Vol. 34, Issue 3, pp. 427-438 (2004)

Vol. 34 Issue 3 pp. 427-438

Electrical substitution cryogenic radiometer based spectral responsivity scale between 250–2500 nm wavelengths

Özcan Bazkir, Farhad Samedov

Keywords

electrical-substitution cryogenic radiometer, electrically calibrated pyroelectric radiometer, trap detector, radiant source, optical power, responsivity

Abstract

This paper presents the spectral responsivity scale between 250–2500 nm developed by the National Metrology Institute of Turkey (UME). For that purpose silicon photodiode based trap detector and electrically calibrated pyroelectric radiometer (ECPR) that were calibrated against primary level absolute electrical substitution cryogenic radiometer (ESCR) were used as transfer standards. Using highly collimated and stabilized (10–5) lasers, absolute optical powers and absolute responsivity of trap detector were measured with an uncertainty of the order of 10–4. In visible (VIS) region responsivity scale was set by means of the models for the reflectance and internal quantum efficiency. In the ultraviolet (UV) and near-infrared (NIR) regions spectrally flat (0.1%) ECPR was used.

Vol. 34
Issue 3
pp. 427-438

0.75 MB

Corresponding address

Optica Applicata
Wrocław University of Science and Technology
Faculty of Fundamental Problems of Technology
Wybrzeże Wyspiańskiego 27
50-370 Wrocław, Poland

Publisher

Wrocław University of Science and Technology
Faculty of Fundamental Problems of Technology
Wybrzeże Wyspiańskiego 27
50-370 Wrocław, Poland

Contact us

  • optica.applicata@pwr.edu.pl
  • +48 71 320 23 93
  • +48 71 328 36 96